The South African National Antiretroviral Treatment Guideline recommends the use of HIV viral load assays for routine monitoring of HIV-1 positive patients on Highly Active Antiretroviral Therapy (HAART). Approved commercial HIV-1 viral load assays are expensive for developing countries where a large number of patients are treated in the public sector. The evaluation of an in-house HIV-1 viral load assay (LUX assay) is described using 458 plasma specimens. Good specificity of the LUX assay was demonstrated using 50 seronegative plasma specimens. A group of 142 HIV-1 positive patients was used to assess the agreement between the LUX assay and the COBAS Amplicor assay. An intra class correlation (ICC) coefficient of 0.85 (CI 95%) indicated good agreement between the assays. The Bland-Altman model showed good agreement between the assays for approximately 87% of the results (mean 0.03 [-1.26; 1.32], CI 95%). In a cohort of 55 patients followed-up longitudinally the LUX assay showed similar declines in viral load to the COBAS Amplicor assay in response to therapy. Viral rebound was detected in 5 patients out of 55 by both assays. Thus, the LUX assay compares well to the gold standard and represents an affordable alternative for high volume testing in resource limited settings.