Quantitation of iptakalim in human plasma by high-performance liquid chromatography-tandem mass spectrometry

J Chromatogr B Analyt Technol Biomed Life Sci. 2007 Nov 15;859(2):256-60. doi: 10.1016/j.jchromb.2007.09.019. Epub 2007 Sep 19.

Abstract

This paper describes a rapid and sensitive analytical method for the quantitation of iptakalim, a novel antihypertensive drug, in human plasma. The method is based on liquid chromatography-tandem mass spectrometry (LC-MS/MS) using sildenafil as internal standard. Sample preparation involved liquid-liquid extraction with dichloromethane-diethyl ether (2:3, v/v) in a basic environment. Chromatography was carried out on an amino column with a mobile phase consisting of acetonitrile-water (55:45, v/v, water containing 0.5% formic acid). Detection employed electrospray ionization (ESI) tandem mass spectrometry in the multiple-reaction-monitoring (MRM) mode. The assay was linear in the concentration range of 0.5-100 ng/ml with a lower limit of quantitation (LLOQ) of 0.5 ng/ml. Intra- and inter-day precision (R.S.D.) were <4.5% and <12.0%, respectively and the accuracy (R.E.) was in the range +/-5%. The method was successfully applied to a single oral dose pharmacokinetic study in human volunteers.

Publication types

  • Validation Study

MeSH terms

  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Propylamines / blood*
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tandem Mass Spectrometry / methods*

Substances

  • N-(1-methylethyl)-1,1,2-trimethylpropylamine
  • Propylamines