Serum free light chains; the need to establish local reference intervals

Ann Clin Biochem. 2007 Nov;44(Pt 6):512-5. doi: 10.1258/000456307782268200.

Abstract

Background: The purpose of this study was to compare The Binding Site serum free light chain assay on two analytical platforms; the Dade Behring BNII and the Olympus AU400. Reference intervals were subsequently established on each analyser.

Methods: In total, 112 serum samples routinely submitted to the laboratory for protein electrophoresis were used for the comparison study. Reference interval data was generated from 126 ostensibly healthy anonymous individuals.

Results: Serum free light chain results on the BNII and AU400 analysers are not directly comparable. The BNII produces results which are significantly higher than the AU400.

Conclusions: Laboratories using The Binding Site serum free light chain assay are strongly advised to establish local reference intervals.

Publication types

  • Comparative Study
  • Evaluation Study

MeSH terms

  • Blood Protein Electrophoresis / standards
  • Clinical Laboratory Techniques / instrumentation
  • Clinical Laboratory Techniques / standards*
  • Humans
  • Immunoassay / methods
  • Immunoglobulin Light Chains / analysis*
  • Immunoglobulin Light Chains / blood*
  • Reference Values

Substances

  • Immunoglobulin Light Chains