High-performance liquid chromatographic determination of iothalamic acid in human plasma and urine

J Chromatogr. 1991 Oct 4;570(2):435-40. doi: 10.1016/0378-4347(91)80550-v.

Abstract

A simple, rapid and sensitive method for the determination of iothalamic acid (IA) in both plasma and urine is reported. After extraction with ethyl acetate, IA was determined by strong anion-exchange high-performance liquid chromatography with ultraviolet detection at 254 nm. The lower limit of detection was 0.5 micrograms/ml. The average recovery was 73 and 57% from plasma and urine, respectively. Linearity was found over the investigated concentration range (up to 500 micrograms/ml for plasma and up to 10.0 mg/ml for urine). The reproducibility of the technique was good (coefficient of variation less than 6%) as was the precision and accuracy (coefficient of variation less than 2.5%). No interference from endogenous substances or any of the common drugs tested was found.

MeSH terms

  • Acetates / chemistry
  • Chromatography, High Pressure Liquid
  • Humans
  • Iothalamic Acid / metabolism*
  • Reproducibility of Results
  • Spectrophotometry, Ultraviolet

Substances

  • Acetates
  • Iothalamic Acid
  • ethyl acetate