Clinical and angiographic results of percutaneous coronary revascularization using a trilayer stainless steel-tantalum-stainless steel phosphorylcholine-coated stent: the TriMaxx trial

Catheter Cardiovasc Interv. 2007 Dec 1;70(7):914-9. doi: 10.1002/ccd.21279.

Abstract

Background: The TriMaxx coronary stent system includes a novel trilayer metal stent having two outer layers of 316L stainless steel and with an inner 0.0007 inches layer of tantalum. This enables creation of a maximally flexible and thin device (0.0029 inches) while still maintaining the requisite strength and radiopacity for effective implantation.

Objectives: The objective of this multi-center, single-arm prospective clinical trial was to assess the safety and performance of the TriMaxx stent for the treatment of single de novo coronary artery lesions.

Methods: One hundred patients with ischemic coronary occlusive disease because of single de novo obstructive lesions of native coronary arteries were treated with 3 x 15 or 3 x 18 mm TriMaxx stents in four hospitals in Brazil and Germany between May of 2004 and September of 2005. An independent core laboratory analyzed the quantitative coronary angiography (QCA) results immediately after stent implantation, and after six months.

Results: The lesion, procedure, and device-deployment success rates were 100, 99, and 100%, respectively. Eighty-eight patients underwent follow-up angiography at 6 +/- 1 months. After six months, 13 (13%) of patients had sustained major adverse cardiac events, including 9.0% that required target lesion revascularization (TLR). The follow-up angiographic studies revealed a binary in-stent restenosis rate of 25% with in-stent late lumen loss of 0.94 +/- 0.57 mm.

Conclusions: These results demonstrate that the TriMaxx stent can be safely deployed for the treatment of single de novo coronary occlusive lesions with six-month clinical and angiographic results rates comparable to historical results using other bare metal stents.

Publication types

  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Brazil
  • Cardiovascular Agents / administration & dosage*
  • Cardiovascular Diseases / diagnostic imaging
  • Cardiovascular Diseases / etiology*
  • Coronary Angiography*
  • Coronary Restenosis / diagnostic imaging
  • Coronary Restenosis / etiology
  • Coronary Stenosis / diagnostic imaging
  • Coronary Stenosis / therapy*
  • Drug-Eluting Stents*
  • Feasibility Studies
  • Female
  • Germany
  • Humans
  • Male
  • Middle Aged
  • Phosphorylcholine / administration & dosage*
  • Prospective Studies
  • Prosthesis Design
  • Severity of Illness Index
  • Stainless Steel*
  • Tantalum*
  • Time Factors
  • Treatment Outcome

Substances

  • Cardiovascular Agents
  • Phosphorylcholine
  • Stainless Steel
  • Tantalum