Studies of whole blood coagulation by oscillatory shear, thromboelastography and free oscillation rheometry

Clin Hemorheol Microcirc. 2008;38(4):267-77.

Abstract

We report studies of the coagulation of samples of whole human blood by oscillatory shear techniques, including Fourier Transform Mechanical Spectroscopy (FTMS). These techniques are used herein to identify the Gel Point of coagulating blood in terms of the Chambon-Winter Gel Point criterion which provides a rheometrical basis for detecting the establishment of an incipient clot. A comparison of the results of FTMS with those obtained from measurements involving a Thromboelastograph (TEG) and a Free Oscillation Rheometer (FOR) indicate that the latter techniques are not capable of detecting the incipient clot, whose establishment occurs several minutes prior to TEG or FOR-based assessments of clot formation time. The results of the present study suggest that FTMS is a useful tool in blood clotting research, being capable of providing a global coagulation profile in addition to detecting the instant of incipient clot formation.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Blood Coagulation / physiology*
  • Hemorheology / instrumentation*
  • Hemorheology / methods*
  • Humans
  • Spectrum Analysis
  • Thrombelastography