Purpose: The aim of the present study was to quantitatively assess the performance of a block matching-based automatic registration algorithm integrated within the commercial treatment planning system designated ISOgray from Dosisoft. The accuracy of the process was evaluated by a phantom study on computed tomography (CT), magnetic resonance (MR) and positron emission tomography (PET) images.
Materials and methods: Two phantoms were used to carry out this study: the cylindrical Jaszczak phantom and the anthropomorphic Liqui-Phil Head Phantom (the Phantom Laboratory), containing fillable spheres. External fiducial markers were used to quantify the accuracy of 41 CT/CT, MR/CT and PET/CT automatic registrations with images of the rotated and tilted phantoms.
Results: The study first showed that a cylindrical phantom was not adapted for the evaluation of the performance of a block matching-based registration software. Secondly, the Liqui-Phil Head Phantom study showed that the algorithm was able to perform automatic registrations of CT/CT and MR/CT images with differences of up to 40 degrees in phantom rotation and of up to 20-30 degrees for PET/CT with accuracy below the image voxel size.
Conclusion: The study showed that the block matching-based automatic registration software under investigation was robust, reliable and yielded very satisfactory results. This phantom-based test can be integrated into a periodical quality assurance process and used for any commissioning of image registration software for radiation therapy.