[Quality control of megavoltage cone beam CT imaging system]

Cancer Radiother. 2008 Dec;12(8):781-7. doi: 10.1016/j.canrad.2008.06.007. Epub 2008 Aug 8.
[Article in French]

Abstract

Purpose: This paper presents the development of a protocol for quality control of a megavoltage cone beam CT imaging system (MVCB) mounted on a Siemens Oncor 6MV linear accelerator.

Materials and methods: Several parameters were controlled on the MVCB system: (1) the initial geometric calibration of the system; (2) the quality of the images (geometric distortion, uniformity, spatial resolution, low contrast resolution) for various protocols; (3) the correspondence between the intensity of voxels and electronic densities; (4) the dose delivered when achieving a MVCB. These tests were done mainly with two cylindrical phantoms specific to the quality control (QC) of a MVCB system, supplied by Siemens, and with the Catphan 600 phantom (The Phantom Laboratory) and Quasar Multipurpose Body phantom (Modus Medical Devices Inc).

Results: The results of the quality control of the images were within the tolerances. The use of the Catphan 600 phantom was inadequate for the QC of MVCB images. These tests also highlighted the need to correct the MVCB images for the "cupping artefact" for dose calculation purpose.

Conclusion: The initial characteristics of the MVCB imaging system were established. Such testing also provided the assessment of the influence of various parameters on the image quality as well as the associated dose delivered during their acquisition, and emphasized the corrections needed to use MVCB images for dose calculation.

Publication types

  • English Abstract

MeSH terms

  • Calibration
  • Humans
  • Particle Accelerators / standards*
  • Phantoms, Imaging
  • Quality Control
  • Radiation Dosage
  • Reproducibility of Results
  • Sensitivity and Specificity
  • Tomography, X-Ray Computed / standards*