Background: This study compared fixation with a titanium one-third tubular plate and one lag-screw vs. fixation with biodegradable plates with one lag-screw applied with two different plate-screw patterns.
Materials and methods: Ten pairs of fibulas were osteotomied, plated (titanium plate with one lag screw vs. absorbable 2/8 plate-screws crossing vs. 0/8 plate-screws crossing the osteotomy gap). and tested in torsion and bending to obtain stiffness and neutral-zone (NZ) data. Tests were performed using 5 load cycles. No load to failure was performed. Biodegradation was simulated by 6 week immersion in phosphate-buffered saline after which the testing protocol was repeated. The specimens were then loaded with 100 N in bending.
Results: Post-implantation, there were no significant differences, in torsion and bending, regarding the NZ or the stiffness, between the 2/8 biodegradable plate and the titanium plate. The 0/8 pattern performed significantly less well in terms of stiffness and NZ in the initial torsion test, and significantly less well in terms of stiffness in the initial bending test. After 6 weeks' immersion, all biodegradable constructs showed a significantly larger NZ and significantly reduced bending and torsional stiffness. When loaded with 100 N, four of the six 0/8 osteosyntheses failed. There were no significant differences between the 2/8 pattern and the titanium plates.
Conclusion: In a model of a Weber-B fracture, the use of a 2/8 biodegradable plate construct initially did not differ statistically to that obtained with a one-third tubular titanium plate. After immersion the 2/8 construct withstood some physiological load.
Clinical relevance: For the fixation of ankle fractures with a biodegradable plate of the type employed in this study, the use of fracture-gap-crossing screws is recommended.