Eur J Cardiothorac Surg. 2009 Jun;35(6):965-9; discussion 969. doi: 10.1016/j.ejcts.2009.02.018. Epub 2009 Mar 26.

Abstract

Objective: Based upon recent developments in transcatheter technology, this study was designed to evaluate the feasibility and haemodynamic performance of transcatheter valve-in-a-ring (VinR) implantation for potentially failed mitral repair using a minimally invasive, transatrial, off-pump approach.

Methods: Adult sheep (54.3+/-3.0 kg) underwent mitral valve repair with a 26 mm complete annuloplasty ring (Physio) using standard conventional techniques. To simulate the redo operation, a transcatheter 23 mm pericardial prosthesis (Edwards Sapien) mounted on a balloon-inflatable steel stent was deployed within the annuloplasty ring. VinR implantation was performed off-pump under rapid pacing in four and on-pump in three animals using an antegrade transatrial approach under fluoroscopic guidance.

Results: Transcatheter VinR implantation was successful in all seven sheep. Mean transvalvular gradient was 4.9+/-0.3 mmHg. VinR function was excellent with no leak in one, good with mild leak in five (trans-stent: four, paravalvular: one) and sufficient with moderate central leak in one animal, respectively. Valve deployment required 10.0+/-0.7 min and all transcatheter prostheses were confirmed in good position on postmortem analysis, without any signs of valve dislocation or embolisation. In an in-vitro model, the minimum force required to dislodge the valve was 32.9+/-5.2N, which was well above the normal estimated forces generated by the left ventricle. One animal was kept alive to assess mid-term outcome and is still well 12 months after the VinR implantation.

Conclusions: Transatrial, transcatheter mitral VinR implantation is feasible using a minimally invasive off-pump approach. VinR implantation is a promising concept for re-operative surgery for selected patients after failed mitral valve repair.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Disease Models, Animal
  • Feasibility Studies
  • Heart Valve Prosthesis Implantation / methods*
  • Heart Valve Prosthesis*
  • Minimally Invasive Surgical Procedures / methods
  • Mitral Valve / diagnostic imaging
  • Mitral Valve / surgery*
  • Mitral Valve Insufficiency / surgery*
  • Prosthesis Design
  • Prosthesis Failure
  • Recurrence
  • Reoperation / methods
  • Sheep
  • Ultrasonography