Effect of different saline chaser volumes and flow rates on intravascular contrast enhancement in CT using a circulation phantom

Eur J Radiol. 2010 Mar;73(3):688-93. doi: 10.1016/j.ejrad.2009.01.008. Epub 2009 Apr 5.

Abstract

Purpose: To evaluate the influence of different saline chaser volumes and different saline chaser flow rates on the intravascular contrast enhancement in MDCT.

Materials and methods: In a physiological flow phantom contrast medium (120 ml, 300 mgI/ml, Ultravist 300) was administered at a flow rate of 6 ml/s followed by different saline chaser volumes (0, 30, 60 and 90 ml) at the same injection rate or followed by a 30-ml saline chaser at different injection rates (2, 4, 6 and 8 ml/s). Serial CT-scans at a level covering the pulmonary artery, the ascending and the descending aorta replica were obtained. Time-enhancement curves were computed and both pulmonary and aortic peak enhancement and peak time were determined.

Results: Compared to contrast medium injection without a saline chaser the pushing with a saline chaser (30, 60, and 90 ml) resulted in a statistically significant increased pulmonary peak enhancement (all p=0.008) and prolonged peak time (p=0.032, p=0.024 and p=0.008, respectively). Highest aortic peak enhancement values were detected for a saline chaser volume of 30 ml. A saline chaser flow rate of 8 ml/s resulted in the highest pulmonary peak enhancement values compared to flow rates of 2, 4 and 6 ml/s (all p=0.008). Aortic peak enhancement showed the highest values for a flow rate of 6 ml/s.

Conclusion: A saline chaser volume of 30 ml and an injection rate of 6 ml/s are sufficient to best improve vascular contrast enhancement in the pulmonary artery and the aorta in MDCT.

MeSH terms

  • Aorta, Thoracic / diagnostic imaging*
  • Contrast Media / administration & dosage*
  • Humans
  • Injections
  • Iohexol / administration & dosage
  • Iohexol / analogs & derivatives*
  • Phantoms, Imaging
  • Pulmonary Artery / diagnostic imaging*
  • Sodium Chloride / administration & dosage*
  • Statistics, Nonparametric
  • Tomography, X-Ray Computed / methods*

Substances

  • Contrast Media
  • Iohexol
  • Sodium Chloride
  • iopromide