Early- and long-term intravascular ultrasound and angiographic findings after bioabsorbable magnesium stent implantation in human coronary arteries

JACC Cardiovasc Interv. 2009 Apr;2(4):312-20. doi: 10.1016/j.jcin.2008.09.015.

Abstract

Objectives: This study aimed to evaluate the degradation rate and long-term vascular responses to the absorbable metal stent (AMS).

Background: The AMS demonstrated feasibility and safety at 4 months in human coronary arteries.

Methods: The PROGRESS-AMS (Clinical Performance and Angiographic Results of Coronary Stenting) was a prospective, multicenter clinical trial of 63 patients with coronary artery disease who underwent AMS implantation. Angiography and intravascular ultrasound (IVUS) were conducted immediately after AMS deployment and at 4 months. Eight patients who did not require repeat revascularization at 4 months underwent late angiographic and IVUS follow-up from 12 to 28 months.

Results: The AMS was well-expanded upon deployment without immediate recoil. The major contributors for restenosis as detected by IVUS at 4 months were: decrease of external elastic membrane volume (42%), extra-stent neointima (13%), and intra-stent neointima (45%). From 4 months to late follow-up, paired IVUS analysis demonstrated complete stent degradation with durability of the 4-month IVUS indexes. The neointima was reduced by 3.6 +/- 5.2 mm(3), with an increase in the stent cross sectional area of 0.5 +/- 1.0 mm(2) (p = NS). The median in-stent minimal lumen diameter was increased from 1.87 to 2.17 mm at long-term follow-up. The median angiographic late loss was reduced from 0.62 to 0.40 mm by quantitative coronary angiography from 4 months to late follow-up.

Conclusions: Intravascular ultrasound imaging supports the safety profile of AMS with degradation at 4 months and maintains durability of the results without any early or late adverse findings. Slower degradation is warranted to provide sufficient radial force to improve long-term patency rates of the AMS.

Publication types

  • Clinical Trial
  • Multicenter Study

MeSH terms

  • Absorbable Implants*
  • Aged
  • Alloys*
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Coronary Angiography*
  • Coronary Artery Disease / diagnostic imaging
  • Coronary Artery Disease / therapy*
  • Coronary Restenosis / diagnostic imaging
  • Coronary Restenosis / etiology
  • Coronary Restenosis / prevention & control*
  • District of Columbia
  • Feasibility Studies
  • Female
  • Humans
  • Magnesium*
  • Male
  • Middle Aged
  • Prospective Studies
  • Prosthesis Design
  • Stents*
  • Time Factors
  • Treatment Outcome
  • Tunica Intima / diagnostic imaging
  • Ultrasonography, Interventional*
  • United States
  • Victoria

Substances

  • Alloys
  • Magnesium