Active two-channel 0.035'' guidewire for interventional cardiovascular MRI

J Magn Reson Imaging. 2009 Aug;30(2):461-5. doi: 10.1002/jmri.21844.

Abstract

Purpose: To develop an "active" (receiver-coil) clinical-grade guidewire with enhanced visibility for magnetic resonance imaging (MRI) and favorable mechanical characteristics for interventional MRI procedures that require conspicuous intravascular instruments distinguishable from surrounding tissues.

Materials and methods: We designed a 0.035-inch guidewire combining two antenna designs on separate channels. A loop antenna visualizes the tip and a dipole antenna visualizes the whole shaft. We compared mechanical characteristics of this guidewire with x-ray alternatives and tested MRI performance at 1.5T in vitro and in vivo in swine.

Results: Images reflected tip position within 0.97 +/- 0.42 mm and afforded whole-shaft visibility under expected conditions without sacrificing device size or handling. We report tip stiffness, torquability, and pushability comparable to commercial interventional guidewires.

Conclusion: Our clinical-grade 0.035-inch active guidewire is conspicuous under MRI and has mechanical performance comparable to x-ray interventional guidewires. This may enable a range of interventional procedures using real-time MRI.

Publication types

  • Research Support, N.I.H., Extramural

MeSH terms

  • Cardiovascular Diseases / therapy*
  • Equipment Design
  • Humans
  • Magnetic Resonance Imaging, Interventional / instrumentation*