Automated precolumn derivatization system for analyzing physiological amino acids by liquid chromatography/mass spectrometry

Biomed Chromatogr. 2010 Jul;24(7):683-91. doi: 10.1002/bmc.1346.

Abstract

An automated method for high-throughput amino acid analysis, using precolumn derivatization high-performance liquid chromatography/electrospray mass spectrometry (HPLC/ESI-MS), was developed and evaluated. The precolumn derivatization step was performed in the reaction port of a home-built auto-sampler system. Amino acids were derivatized with 3-aminopyridyl-N-hydroxysuccinimidyl carbamate, and a 3 microm Wakosil-II 3C8-100HG column (100 x 2.1 mm i.d.) was used for separation. To achieve a 13 min cycle for each sample, the derivatization and separation steps were performed in parallel. The results of the method evaluation, including the linearity, and the intra- and inter-precision, were sufficient to measure physiological amino acids in human plasma samples. The relative standard deviations of typical amino acids in actual human plasma samples were below 10%.

Publication types

  • Evaluation Study

MeSH terms

  • Amino Acids / analysis*
  • Amino Acids / blood
  • Amino Acids / metabolism
  • Chromatography, High Pressure Liquid / instrumentation
  • Chromatography, High Pressure Liquid / methods*
  • Humans
  • Spectrometry, Mass, Electrospray Ionization / instrumentation
  • Spectrometry, Mass, Electrospray Ionization / methods*

Substances

  • Amino Acids