A rapid and simple liquid chromatography-fluorescence detection (LC-FD) method was developed and validated for the simultaneous quantification of irinotecan (CPT11) and SN38 in sheep plasma. Camptothecin (CPT) was used as the internal standard. A single step protein precipitation with acetonitrile was used for sample preparation. The separation was achieved using a 5 microm C18 column (250 mm x 4.5 mm, 5 microm) with a mobile phase composed of 36 mM sodium dihydrogen phosphate dehydrate and 4 mM sodium 1 heptane sulfonate-acetonitrile (72:28), the pH of the mobile phase was adjusted to 3. The flow rate was 1.45 mL/min and the fluorescence detection was operated at 355/515 nm (excitation/emission wavelengths). The run time was 13 min. The method was validated with respect to selectivity, extraction recovery, linearity, intra- and inter-day precision and accuracy, limit of quantification and stability. The method has a limit of quantification of 5 ng/mL for both CPT11 and SN38. The assay was linear over concentrations ranging from 5 to 5000 ng/mL and to 240 ng/mL for CPT11 and SN38, respectively. This method was used successfully to perform plasma pharmacokinetic studies of CPT11 after pulmonary artery embolization (PACE) in a sheep model. It was also validated for CPT11 and SN38 analysis in sheep lymph and human plasma.
2010. Published by Elsevier B.V.