Effects of lesion positioning on digital magnification mammography performance

Acad Radiol. 2010 Jun;17(6):791-4. doi: 10.1016/j.acra.2010.02.007.

Abstract

Rationale and objectives: We undertook this study to determine whether differences in detector-lesion distance resulted in appreciable effects on digital magnification mammography performance as measured using the American College of Radiology (ACR) mammography phantom and a line pair test pattern.

Materials and methods: Images of the standard 42-mm thick standard ACR mammography phantom with a wax insert on one side containing simulated fibers, calcifications, and masses were obtained on a Senographe Essential digital mammography system with the phantom in upright and inverted positions. The process was repeated with a line pair test pattern for measuring resolution. All images were obtained in contact mode, and with 1.5x and 1.8x magnification, and evaluated on a GE PACS monitor.

Results: Overall, changing lesion-detector distance using standard versus inverted positioning did not appreciably increase the number of objects seen on the ACR phantom under all modes. No greater than one line pair difference was seen in standard versus inverted positioning. At 1.8x magnification mode, no difference was detected in line pair resolution with a change in positioning.

Conclusion: Differences in lesion-detector distance as modeled using both the ACR mammography phantom and a line pair test pattern did not make an appreciable difference in digital magnification mammography performance.

MeSH terms

  • Algorithms*
  • Artifacts*
  • Breast Neoplasms / diagnostic imaging*
  • Female
  • Humans
  • Phantoms, Imaging
  • Radiographic Image Enhancement / instrumentation
  • Radiographic Image Enhancement / methods*
  • Radiographic Image Interpretation, Computer-Assisted / methods*
  • Reproducibility of Results
  • Sensitivity and Specificity