Objective: To evaluate the analytical performance of the Nova StatSensor creatinine meter.
Design and methods: Imprecision, linearity, patient correlation and interference studies were completed.
Results: Total imprecision was 4.5-9.1%CV and a linear measurement range of 93-863 micromol/L (1.05-9.76 mg/dL) was verified. Whole blood creatinine results correlated well with laboratory plasma measurements (R(2)=0.9328) but exhibited a negative proportional bias. In vitro, high levels of creatine and urea falsely elevated creatinine measurement.
Conclusions: The creatinine meter provides reliable measurement across a clinically relevant range and has potential use in point-of-care testing.