A prototype chopper for synchrotron time-resolved crystallographic measurements

Rev Sci Instrum. 2010 Apr;81(4):043905. doi: 10.1063/1.3358939.

Abstract

A mechanical x-ray chopper has been designed to perform microsecond time-resolved crystallographic studies at the DIAMOND synchrotron I19 beamline. It consists of two asymmetric absorbers rotating synchronously at frequencies from 0 to 50 Hz in the same direction around a rotation axis that is parallel to the x-ray beam. The duration of the x-ray pulses produced by the chopper is determined by the relative phase between the two blades, which can be adjusted. The chopper system presented in this paper offers a time resolution suitable for conducting in situ experiments that afford the crystal structure of materials while in their transient (>10 micros) photoactivated excited states.