Evaluation of the TEST 1 erythrocyte sedimentation rate system and intra- and inter-laboratory quality control using new latex control materials

Clin Chem Lab Med. 2010 Jul;48(7):1043-8. doi: 10.1515/CCLM.2010.162.

Abstract

Background: The erythrocyte sedimentation rate (ESR) test has been considered to be a simple procedure, not requiring quality control (QC). However, QC is essential for accuracy and precision. We evaluated the TEST 1 ESR system and performed QC procedures using newly developed latex control materials in three hospitals.

Methods: Using tripotassium ethylenediaminetetraacetic acid blood samples (n=184), we compared TEST 1 ESR values with Westergren ESR data and evaluated intra-assay precision. Three levels of latex control materials were used to assess inter-assay precision. Reference range assessment was done using samples from 220 healthy individuals. Inter-laboratory QC with latex control materials in three hospitals was performed.

Results: Correlation between TEST 1 ESR and Westergren ESR results was good (p<0.001). Intra-assay precision [coefficients of variation (CV) 6.6%-21.7%] with patient samples and inter-assay precision (CV 0.0%-6.8%) with latex control materials were satisfactory. The reference ranges of 2-10 mm/h for males and 2-19 mm/h for females were established. Inter-laboratory QC data with latex control materials in three hospitals demonstrated good accuracy and satisfactory precision (CV 0.0%-14.4%).

Conclusions: Our results demonstrate that the TEST 1 QC is reliable and the latex control materials are valuable for inter-laboratory proficiency testing.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adolescent
  • Adult
  • Aged
  • Blood Sedimentation*
  • Female
  • Humans
  • Laboratories, Hospital / standards*
  • Latex / chemistry*
  • Male
  • Middle Aged
  • Quality Control
  • Reference Values

Substances

  • Latex