Image-derived input function from cardiac gated maximum a posteriori reconstructed PET images in mice

Mol Imaging Biol. 2011 Apr;13(2):342-7. doi: 10.1007/s11307-010-0347-4.

Abstract

Purpose: The purpose of this study was to determine if accurate image-derived input functions (IDIF) can be measured from cardiac gated positron emission tomography (PET) images reconstructed using ordered subset expectation maximization-maximum a posteriori (OSEM-MAP) without further correction.

Procedures: IDIFs from the left ventricle were measured from cardiac gated PET images reconstructed using OSEM-MAP with computed tomography (CT)-based attenuation correction for five C57/BL6 mice. The accuracy of the IDIF was tested against blood samples using Bland-Altman analysis.

Results: Image-derived blood radioactivity concentration values were not significantly different from sampled blood values at two late time points as determined by a paired t test (P = 0.97). Bland-Altman analysis revealed a mean difference of 0.06 μCi/ml (1%). Using kinetic analysis, the mean myocardial 2-deoxy-2-[(18)F]fluoro-D-glucose uptake rate constant based on the IDIF was comparable to values reported in the literature based on physical blood sampling.

Conclusions: Accurate IDIFs can be obtained non-invasively. Although reconstruction times are increased, no further spillover corrections are necessary for IDIFs derived from gated, OSEM-MAP reconstructed images with attenuation correction.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Cardiac-Gated Imaging Techniques / methods*
  • Fluorodeoxyglucose F18
  • Heart / diagnostic imaging
  • Heart / physiology*
  • Image Enhancement / methods*
  • Mice
  • Mice, Inbred C57BL
  • Phantoms, Imaging
  • Positron-Emission Tomography / methods*

Substances

  • Fluorodeoxyglucose F18