Mechanical comparison of meniscal repair devices with mattress suture devices in vitro

Knee Surg Sports Traumatol Arthrosc. 2010 Nov;18(11):1594-8. doi: 10.1007/s00167-010-1188-z. Epub 2010 Jun 9.

Abstract

We report the load to failure in tensile testing of the MaxFire™ meniscal repair system (Biomet Inc, Warsaw, IN) and compare it to other current meniscal repair devices and mattress suture techniques. After creating a longitudinal tear in 42 one-year-old bovine menisci, 7 specimen groups defined by the meniscal repair device, suture, and/or mattress technique used for meniscal repair were randomly established: (Group 1: Fiberwire™ vertical mattress (VM), Group 2: Fiberwire™ horizontal mattress (HM), Group 3: FasT-Fix™ VM, Group 4: FasT-Fix™ HM, Group 5: RapidLoc™, Group 6: MaxFire™ VM, Group 7: MaxFire™ HM). After completing the repairs, the meniscal specimens were cyclically pre-loaded before load to failure testing was performed. The mean load to failure for each group was: Fiberwire VM (185 ± 41 N), Fiberwire HM (183 ± 36 N), FasT-Fix VM (125 ± 8 N), FasT-Fix HM (107 ± 29 N), RapidLoc (70 ± 12 N), MaxFire VM (145 ± 44 N), MaxFire HM (139 ± 50 N). An analysis of variance demonstrated a significant difference in the mean load to failure (F = 8.31 P < 0.01). Statistically significant differences were seen between both Fiberwire groups verses FasT-Fix HM and Rapid-Loc (P < 0.05). Three modes of failure were observed: suture breakage (17/42, 40.5%), tissue failure (18/42, 42.9%), and knot failure (7/42, 16.7%). 2-0 Fiberwire™ VM and HM repairs had the highest load to failure of all groups tested. The load to failure for the MaxFire™ meniscal repair system is comparable to other available all-inside meniscal repair systems.

Publication types

  • Comparative Study

MeSH terms

  • Analysis of Variance
  • Animals
  • Biomechanical Phenomena
  • Cattle
  • Equipment Design
  • Equipment Failure Analysis
  • In Vitro Techniques
  • Materials Testing
  • Menisci, Tibial / surgery*
  • Orthopedic Fixation Devices
  • Reference Values
  • Stress, Mechanical*
  • Suture Techniques
  • Sutures
  • Tensile Strength