Human cardiovascular effects of a new generation conducted electrical weapon

Forensic Sci Int. 2011 Jan 30;204(1-3):50-7. doi: 10.1016/j.forsciint.2010.05.003.

Abstract

Objectives: The conducted electrical weapon (CEW) is used by law enforcement to control potentially violent people. Much of the research in CEW safety has focused on the TASER X26, which uses a single deployment cartridge. New Generation CEW (NGCEW) technology has been developed that uses a different circuit and multiple cartridges that can be simultaneously deployed. The objective of this study is to examine the cardiovascular effects of the NGCEW in different deployment possibilities.

Methods: This was a prospective study of human subjects during NGCEW training courses. Subjects received a NGCEW probe deployment to the frontal torso in 1 of 3 configurations: 2, 3,or 4 embedded probes and then underwent a 10-s exposure. Before and after vital signs, electrocardiograms (ECGs), and serum troponin I values were obtained. Real-time echocardiography was utilized before, during and after the exposure to evaluate heart rate and rhythm.

Results: Initially, a 1st version NGCEW (NGCEWv1) that was in the final stages of manufacturer verification was used at the training courses. It had not been publicly released. During a NGCEWv1 exposure with 2 probes, there was an apparent brief episode of cardiac capture. Testing was halted and the manufacturer was notified. The device was redesigned and the study continued when a redesigned, 2nd version (NGCEWv2) was used. The NGCEW1 was studied in 8 subjects. The NGCEWv2 was studied in 45 subjects with no evidence of cardiac capture. There were no important post-exposure vital sign, troponin I or ECG changes found in any volunteers.

Conclusions: An apparent brief myocardial capture event occurred with the NGCEWv1. This device was not released and was redesigned. The NGCEWv2 appears to exhibit a reasonable degree of cardiac safety with frontal torso exposures and multiple probe combination configurations.

Publication types

  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Arrhythmias, Cardiac / etiology*
  • Blood Pressure
  • Echocardiography*
  • Electric Stimulation / instrumentation*
  • Electrocardiography
  • Equipment Design
  • Female
  • Heart Rate
  • Humans
  • Law Enforcement
  • Male
  • Middle Aged
  • Oxygen / blood
  • Prospective Studies
  • Troponin I / blood
  • Weapons*

Substances

  • Troponin I
  • Oxygen