Test-re-test reproducibility of Doppler echocardiography for assessment of electromechanical dyssynchrony: implications for heart failure clinic

J Cardiol. 2010 Nov;56(3):271-9. doi: 10.1016/j.jjcc.2010.06.002. Epub 2010 Jul 16.

Abstract

Background: Reproducibility of Doppler echocardiography for assessment of inter-ventricular and intra-left ventricular (LV) dyssynchrony, and its clinical implications, have not been established.

Methods: Twenty-eight subjects (heart failure stages A-C, 61% with QRS ≥ 120 ms, ejection fraction (EF) ≤ 35%) underwent two consecutive echo-studies within 24h to evaluate test-re-test reproducibility of inter-ventricular electromechanical delay (VV delay, by traditional pulsed-Doppler), and intra-LV electromechanical delay between opposite LV walls by color-coded Doppler tissue-velocity (COLOR-DTI), and by pulsed-Doppler tissue spectrum (PW-DTI). Reproducibility of LV internal diastolic diameter (LVIDD) and of EF (by Simpson's method) assessments was evaluated contextually for reference.

Results: Intra-study and inter-study reproducibility of inter-ventricular and intra-LV electromechanical dyssynchrony was in general good, and comparable to the reproducibility of LVIDD and EF assessments. Between-study reproducibility of PW-TDI method was fair, but showed poor agreement with COLOR-TDI method. In repeated studies, agreement of significant electromechanical delay by COLOR-TDI was comparable to the agreement of EF ≤ 35%. In the 5 patients who had simultaneously large QRS, EF ≤ 35%, and significant inter- and intra-ventricular dyssynchrony at study #1, 3 had EF 36-40% and 1 showed no significant dyssynchrony by study #2.

Conclusion: In serial echocardiographic studies, Doppler echocardiography showed a good test-re-test reproducibility for the identification of significant electromechanical delay. Planimetry for EF assessment was a source of variability as relevant as Doppler echocardiography, but COLOR-DTI may add meaningful and reproducible information to QRS duration for cardiac-resynchronization therapy.

MeSH terms

  • Adult
  • Aged
  • Cardiac Resynchronization Therapy
  • Echocardiography, Doppler / methods*
  • Female
  • Heart Failure / diagnostic imaging*
  • Heart Failure / physiopathology
  • Heart Failure / therapy
  • Humans
  • Male
  • Middle Aged
  • Reproducibility of Results
  • Stroke Volume
  • Ventricular Function, Left
  • Ventricular Function, Right