A liquid chromatography-tandem mass spectrometry method for the simultaneous quantification of escin Ia and escin Ib in human plasma: application to a pharmacokinetic study after intravenous administration

Biomed Chromatogr. 2010 Dec;24(12):1309-15. doi: 10.1002/bmc.1441.

Abstract

A rapid and sensitive liquid chromatography-tandem mass spectrometry (LC-MS/MS) method was developed and validated for simultaneous quantification of escin Ia and escin Ib in human plasma. After a solid-phase extraction (SPE), the analytes were separated on a Zorbax Extend C(18) column by isocratic elution with a mobile phase of methanol-acetonitrile-10 mm ammonium acetate (27:27:46, v/v/v) at a flow rate of 1.0 mL/min and analyzed by mass spectrometry in the positive ion multiple reaction monitoring mode. The precursor to product ion transitions of m/z 1131.8 → 807.6 was used to quantify escin Ia and escin Ib. Good linearity was achieved over a wide range of 2.00-900 ng/mL for escin Ia and 1.50-662 ng/mL for escin Ib. The intra- and inter-day precisions (as relative standard deviation) were less than 11% for each QC level of escin Ia and escin Ib. The accuracies (as relative error) were within ±5.27% for escin Ia and within ±4.07% for escin Ib. The method was successfully employed in a pharmacokinetic study after a single intravenous infusion administration of sodium aescinate injection containing 10 mg escin to each of the 10 healthy volunteers.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Chromatography, Liquid / methods*
  • Escin / administration & dosage
  • Escin / blood*
  • Escin / isolation & purification
  • Escin / pharmacokinetics
  • Humans
  • Injections, Intravenous
  • Solid Phase Extraction
  • Tandem Mass Spectrometry / methods*

Substances

  • escin Ib
  • Escin
  • escin Ia