In vivo sodium imaging of human patellar cartilage with a 3D cones sequence at 3 T and 7 T

J Magn Reson Imaging. 2010 Aug;32(2):446-51. doi: 10.1002/jmri.22191.

Abstract

Purpose: To compare signal-to-noise ratios (SNRs) and T*(2) maps at 3 T and 7 T using 3D cones from in vivo sodium images of the human knee.

Materials and methods: Sodium concentration has been shown to correlate with glycosaminoglycan content of cartilage and is a possible biomarker of osteoarthritis. Using a 3D cones trajectory, 17 subjects were scanned at 3 T and 12 at 7 T using custom-made sodium-only and dual-tuned sodium/proton surface coils, at a standard resolution (1.3 x 1.3 x 4.0 mm(3)) and a high resolution (1.0 x 1.0 x 2.0 mm(3)). We measured the SNR of the images and the T*(2) of cartilage at both 3 T and 7 T.

Results: The average normalized SNR values of standard-resolution images were 27.1 and 11.3 at 7 T and 3 T. At high resolution, these average SNR values were 16.5 and 7.3. Image quality was sufficient to show spatial variations of sodium content. The average T*(2) of cartilage was measured as 13.2 +/- 1.5 msec at 7 T and 15.5 +/- 1.3 msec at 3 T.

Conclusion: We acquired sodium images of patellar cartilage at 3 T and 7 T in under 26 minutes using 3D cones with high resolution and acceptable SNR. The SNR improvement at 7 T over 3 T was within the expected range based on the increase in field strength. The measured T*(2) values were also consistent with previously published values.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Adult
  • Biomarkers / metabolism
  • Cartilage / pathology
  • Female
  • Glycosaminoglycans / chemistry
  • Humans
  • Image Processing, Computer-Assisted
  • Magnetic Resonance Imaging / methods*
  • Male
  • Osteoarthritis / diagnosis
  • Osteoarthritis / pathology*
  • Patella / pathology*
  • Reproducibility of Results
  • Sodium / chemistry*
  • Time Factors

Substances

  • Biomarkers
  • Glycosaminoglycans
  • Sodium