A disposable electrochemical immunosensor using screen-printed carbon electrodes (SPCEs) and protein A-functionalized magnetic beads (MBs) was developed for the determination of testosterone. Anti-testosterone was immobilized onto MBs and a direct competitive immunoassay involving testosterone labeled with peroxidase (HRP) was performed. The resulting conjugate was trapped on the SPCE with a small magnet. Testosterone determination was carried out by amperometry at -0.2V upon H2O2 additions using hydroquinone (HQ) as the redox mediator. The experimental variables involved in the immunosensor response to testosterone were evaluated. Under the optimized conditions, a calibration plot for testosterone was obtained with a linear range between 5.0×10(-3) and 50 ng/mL (r=0.995). The detection limit was 1.7 pg/mL and the EC50 was 0.25±0.04 ng/mL. These characteristics are notably better than those achieved with other reported immunosensors. Furthermore, anti-testosterone/MBs conjugates were shown to be stable for at least 25 days. A good selectivity was also found against other steroid hormones. The usefulness of the immunosensor was demonstrated by analyzing human serum spiked with 1 and 10 ng/mL testosterone.
Copyright © 2010 Elsevier B.V. All rights reserved.