We prospectively evaluated acetabular cup placement in total hip arthroplasty with an imageless computer navigation system or using conventional manual technique. The achieved cup orientation in the manual group had substantially larger variances and greater placement error than the navigation cases. The use of navigation was abandoned in 3 cases because of excessive pelvic tilt and unreliable registration of the pelvis. Computer navigation system helped improve the accuracy of the acetabular cup placement for total hip arthroplasty in this study. The variation between the intraoperative navigation readings and the computed tomographic values suggests that relying on palpation of bony landmarks through drapes and tissue is a limitation of this method. Further, the variation in pelvic tilt has an effect on cup placement that requires further studies.
Copyright © 2011 Elsevier Inc. All rights reserved.