Thrombosis of second-generation drug-eluting stents in real practice results from the multicenter Spanish registry ESTROFA-2 (Estudio Español Sobre Trombosis de Stents Farmacoactivos de Segunda Generacion-2)

JACC Cardiovasc Interv. 2010 Sep;3(9):911-9. doi: 10.1016/j.jcin.2010.06.010.

Abstract

Objectives: This study sought to evaluate second-generation drug-eluting stent (DES) thrombosis in clinical practice.

Background: First-generation DES are associated with a significant incidence of late thrombosis. There is paucity of data regarding real practice late thrombosis incidence and predictors with second-generation DES, zotarolimus-eluting stent (ZES), and everolimus-eluting stents (EES).

Methods: A prospective, large-scale, non-industry-linked multicenter registry was designed. Complete clinical-procedural data and systematic follow-up of all patients treated with these stents was reported in a dedicated registry supported by the Spanish Working Group on Interventional Cardiology.

Results: From 2005 to 2008, 4,768 patients were included in 34 centers: 2,549 treated with ZES, and 2,219 with EES. The cumulative incidence of definite/probable thrombosis for ZES was 1.3% at 1 year and 1.7% at 2 years and for EES 1.4% at 1 year and 1.7% at 2 years (p = 0.8). The increment of definite thrombosis between the first and second year was 0.2% and 0.25%, respectively. In a propensity score analysis, the incidence remained very similar. Ejection fraction (adjusted hazard ratio [HR]: 0.97; 95% confidence interval [CI]: 0.95 to -0.99; p = 0.008), stent diameter (adjusted HR: 0.37; 95% CI: 0.17to 0.81; p = 0.01) and bifurcations (adjusted HR: 2.1; 95% CI: 1.14 to 3.7; p = 0.02) emerged as independent predictors of thrombosis. In the subgroup of patients with bifurcations, the use of ZES was independently associated with a higher thrombosis rate (adjusted HR: 4; 95% CI: 1.1 to 13; p = 0.03).

Conclusions: In a real practice setting, the incidence of thrombosis at 2 years with ZES and EES was low and quite similar. The incidence of very late thrombosis resulted lower than was reported in registries of first-generation DES. In the subset of bifurcations, the use of ZES significantly increased the risk of thrombosis.

Publication types

  • Comparative Study
  • Multicenter Study

MeSH terms

  • Aged
  • Coronary Restenosis / epidemiology*
  • Coronary Restenosis / etiology
  • Coronary Thrombosis / epidemiology*
  • Coronary Thrombosis / etiology
  • Drug-Eluting Stents / adverse effects*
  • Female
  • Follow-Up Studies
  • Humans
  • Incidence
  • Male
  • Prospective Studies
  • Prosthesis Failure
  • Registries*
  • Risk Factors
  • Spain / epidemiology