Multicenter evaluation of a new automated electrochemiluminescence immunoassay for the quantification of testosterone compared to liquid chromatography tandem mass spectrometry

Clin Biochem. 2011 Feb;44(2-3):264-7. doi: 10.1016/j.clinbiochem.2010.09.024. Epub 2010 Oct 12.

Abstract

Objectives: A new electrochemiluminescence testosterone immunoassay (Elecsys® Testo II) was evaluated.

Design and methods: Within-run precision, total precision, and method comparisons to the former immunoassay and LC-MS/MS were investigated. Reference intervals were established for adults and children.

Results: Imprecision ranged from 0.9% to 8.5%. Method comparisons revealed slopes from 0.68 to 1.32 and coefficients of correlation from 0.893 to 0.991.

Conclusions: Improved agreement with LC-MS/MS was demonstrated, especially for female patients. The assay may be a suitable alternative for laboratories with no access to MS techniques.