A hardened gated x-ray imaging diagnostic for inertial confinement fusion experiments at the National Ignition Facility

Rev Sci Instrum. 2010 Oct;81(10):10E539. doi: 10.1063/1.3478897.

Abstract

A gated x-ray detector is under development for use at the National Ignition Facility that is intended to provide plasma emission images in the presence of neutron yields up to 10(15) expected during inertial confinement fusion experiments with layered cryogenic targets. These images are expected to provide valuable time-resolved measurements of core and fuel symmetries. Additional capabilities of this instrument will include the ability to make spatially resolved electron temperature measurements. A description of this instrument and its operation is given with emphasis on features that differentiate it from previous designs.