Biomechanical evaluation of a new device for internal fixation of femoral neck fractures

Artif Cells Blood Substit Immobil Biotechnol. 2011 Aug;39(4):252-8. doi: 10.3109/10731199.2010.538404. Epub 2010 Nov 30.

Abstract

This study aimed to compare the fixation stability of a new device (SCCAB) for femoral neck fractures and the cancellous lag screws. Standardized femoral neck osteotomies were created in seven pairs of cadaver femurs and stabilized with either SCCAB or cancellous lag screws. The specimens were subjected to incremental axial loading to 1300 N and cyclic loading at 1300 N for 104 cycles. No significant differences in downward femoral head displacement, change in the superior osteotomy gapping, and load to failure were observed between the two groups. SCCAB may be potentially useful to treat femoral neck fractures.