Background: There is increasing interest in measuring the soluble forms of carbonic anhydrase IX (CA IX) in blood as a marker of hypoxia for prognostic purposes or for predictive use in therapeutic trials in various cancers. Following our initial observations of marked differences in the measured concentrations of CA IX in EDTA plasma versus serum, we sought to investigate these further in order to determine their effects on results in published studies and to ensure accurate measurement in future studies.
Methods: Serum and EDTA plasma samples from healthy controls and patients with renal cancer were used in the validation of two commercially available enzyme-linked immunosorbent assays (ELISAs) for CA IX with examination of recovery, parallelism and specificity and comparison of paired plasma and serum.
Results: Successful validation of one of the ELISAs was not achieved with particular problems with parallelism and marked differences in measured CA IX concentrations between EDTA plasma and serum. This appeared to be due to a metal ion-dependent epitope on CA IX recognized by the detection antibody in this assay. The other commercially available ELISA examined was successfully validated and showed no difference in CA IX between EDTA plasma and serum.
Conclusions: These results have important consequences for published studies using this assay where the conclusions drawn from the measurements made may be invalid. This study highlights the need for stringent validation of commercially available assays, including examination of various sample types, before use in research studies.