[Pullout test in expansive unilateral open-door laminoplasty of the cervical spine with OsteoMed M3 plate and screws: the screw orientation matters]

Nan Fang Yi Ke Da Xue Xue Bao. 2011 May;31(5):911-3.
[Article in Chinese]

Abstract

Objective: To study the impact of screw orientation on the pullout strength of OsteoMed M3 titanium screws in expansive unilateral open-door laminoplasty of the cervical spine.

Methods: Six fresh human cervical spine specimens were randomly numbered and OsteoMed M3 plate and screws were used for an expansive unilateral open-door laminoplasty. The screws were inserted in the lateral mass at different extraversion angles (0°, 30° and 45°). The maximum pullout strength was tested on the ElectroForce material testing machine.

Results: The maximum pullout strength was 81.60∓7.33 N, 150.05∓15.57 N, and 160.08∓17.77 N in extraversion angle 0°, 30°, and 45° groups, respectively. The maximum pullout strength was significantly less in extraversion angle 0° group than in 30° and 45° groups (P<0.05), but similar in the latter two groups.

Conclusion: The pullout strength of the screws inserted at an extraversion angle over 30° provides stronger fixation than an angle of 0° in the unilateral open-door laminoplasty using OsteoMed M3 titanium plate and screws.

Publication types

  • English Abstract
  • Research Support, U.S. Gov't, Non-P.H.S.

MeSH terms

  • Adult
  • Biomechanical Phenomena
  • Bone Plates
  • Bone Screws
  • Cervical Vertebrae / surgery
  • Cervicoplasty / instrumentation*
  • Device Removal
  • Fracture Fixation, Internal / instrumentation*
  • Humans
  • Internal Fixators
  • Male
  • Materials Testing
  • Young Adult