Comparison of two electromyographical endotracheal tube systems for intraoperative recurrent laryngeal nerve monitoring: reliability and side effects

Langenbecks Arch Surg. 2011 Dec;396(8):1173-9. doi: 10.1007/s00423-011-0819-2. Epub 2011 Jun 29.

Abstract

Background: Recurrent laryngeal nerve (RLN) monitoring systems should be reliable and safe. Monitoring via electromyographical systems on an endotracheal tube (ETT) is widely spread. The MagStim™ system consists of an adhesive electrode to be fixed on an endotracheal tube. The Xomed™ endotracheal tube provides integrated electrodes. Reliability and side effects had never been compared. As both systems have very different morphological properties, we hypothesized that there might be differences in reliability and the incidence of side effects.

Methods: In a retrospective quality management analysis of 118 patients (MagStim™ electrode, 57 patients; Xomed™ ETT, 61 patients), we compared laryngeal side effects according to the Chilla score and detection rate of the RLN.

Results: Both systems had comparable detection rates of the RLN above 95%. Both electrode systems seemed to have similar reliability. Difficulties to detect the nerve were observed in seven patients (four with MagStim, three with Xomed). In the group with the Xomed™ ETT, significantly less mild laryngeal side effects were observed.

Conclusion: Both MagStim™ and Xomed™ ETT were reliable in detecting the RLN. The Xomed™ ETT, however, might cause milder laryngeal side effects compared with the MagStim™ electrode.

Publication types

  • Comparative Study

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Analysis of Variance
  • Cohort Studies
  • Electromyography / instrumentation*
  • Electromyography / methods
  • Female
  • Humans
  • Intraoperative Complications / prevention & control
  • Intubation, Intratracheal / instrumentation*
  • Intubation, Intratracheal / methods*
  • Male
  • Middle Aged
  • Monitoring, Intraoperative / instrumentation*
  • Monitoring, Intraoperative / methods
  • Parathyroidectomy / adverse effects
  • Parathyroidectomy / methods
  • Recurrent Laryngeal Nerve / physiology*
  • Reproducibility of Results
  • Retrospective Studies
  • Risk Assessment
  • Sensitivity and Specificity
  • Statistics, Nonparametric
  • Thyroidectomy / adverse effects
  • Thyroidectomy / methods
  • Treatment Outcome
  • Young Adult