Radiation doses to paediatric patients and comforters undergoing chest X rays

Radiat Prot Dosimetry. 2011 Sep;147(1-2):171-5. doi: 10.1093/rpd/ncr295. Epub 2011 Jul 9.

Abstract

Pneumonia is an important cause of hospital admission among children in the developed world and it is estimated to be responsible for 3-18 % of all paediatric admissions. Chest X ray is an important examination for pneumonia diagnosis and for evaluation of complications. This study aims to determine the entrance surface dose (ESD), organ, effective doses and propose a local diagnostic reference level. The study was carried out at the university hospital of Larissa, Greece. Patients were divided into three groups: organ and effective doses were estimated using National Radiological Protection Board software. The ESD was determined by thermoluminescent dosemeters for 132 children and 76 comforters. The average ESD value was 55 ± 8 µGy. The effective dose for patients was 11.2 ± 5 µSv. The mean radiation dose for comforter is 22 ± 3 µGy. The radiation dose to the patients is well within dose constraint, in the light of the current practice.

MeSH terms

  • Child
  • Child, Preschool
  • Humans
  • Infant
  • Infant, Newborn
  • Pediatrics*
  • Pneumonia / diagnosis*
  • Radiation Dosage*
  • Radiation Monitoring / standards*
  • Radiation Protection / standards*
  • Radiography, Thoracic*
  • Thermoluminescent Dosimetry*