Automated determination of rifampicin in plasma samples by in-tube solid-phase microextraction coupled with liquid chromatography

J Chromatogr B Analyt Technol Biomed Life Sci. 2011 Aug 15;879(24):2454-8. doi: 10.1016/j.jchromb.2011.06.041. Epub 2011 Jul 6.

Abstract

A sensitive and automated method is described for determination of rifampicin in plasma samples for therapeutic drug monitoring by in-tube solid-phase microextraction coupled with liquid chromatography (in-tube SPME/LC). Important factors in the optimization of in-tube SPME are discussed, such as coating type, sample pH, sample draw/eject volume, number of draw/eject cycles, and draw/eject flow rate. Analyte pre-concentrated in the polyethylene glycol phase was directly transferred to the liquid chromatographic column by percolation of the mobile phase, without carryover. The method was linear over the 0.1-100 μg/mL range, with a linear coefficient value (r(2)) of 0.998. The inter-assay precision presented coefficient of variation ≤ 1.7%. The effectiveness and practicability of the proposed method are proven by analysis of plasma samples from ageing patients undergoing therapy with rifampicin.

Publication types

  • Evaluation Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Automation
  • Chromatography, Liquid / methods*
  • Humans
  • Rifampin / blood*
  • Rifampin / isolation & purification*
  • Solid Phase Microextraction / methods*

Substances

  • Rifampin