Objectives: This study evaluated the effect of post-polymerization heat-treatments on degree of conversion (DC), residual methyl methacrylate concentration (MMA(r)) and in vitro cytotoxicity of autopolymerizing acrylic repair resin.
Methods: A total of 336 specimens were prepared by bench- and hydroflask-curing and subjected to post-polymerization heat-treatments: a) water immersion at 60°C for 30 min, b) microwaving at 500 W for 3 min, c) combined use of water immersion and microwaving d) no treatment (as control). Specimens were eluted in cell culture medium for 1, 2, 5 and 7 days. DC and MMA(r) in eluates were measured by FTIR spectrometry and HPLC, respectively. In vitro cytotoxicity of eluates on L-929 fibroblasts was determined by XTT assay. Data were statistically analyzed with Dunn's multiple comparison and Pearson correlation tests (p≤0.05).
Results: DC was highest (99.9%) in bench- and hydroflask-cured groups which were subjected to water immersion. At all elution periods, MMA(r) was detected in eluates of all treatment groups and were higher in bench-cured groups than hydro-flask cured groups. Cell proliferation values indicated slightly cytotoxic effect throughout 7 days; regardless of the curing method or post-polymerization treatment. The correlation between MMA(r) and cell proliferation was negative after elution of 1, 2, 5 days and was only statistically significant (p<0.05) at 5 days. At elution of 7 days, the correlation was positive with no significance.
Significance: Post-polymerization heat-treatment of autopolymerizing acrylic repair resin by immersion in water at 60°C for 30 min is clinically recommended to improve the DC while reducing the leaching residual MMA.
Copyright © 2011 Academy of Dental Materials. Published by Elsevier Ltd. All rights reserved.