Objectives: Computer-aided joint space analysis (CAJSA) is a recently developed, semi-automated tool to aid in the measurement of joint space margins based on hand radiographs. The objective of this study is to verify the potential effect of hand rotation during X-ray imaging on the measurement of joint space width using CAJSA and to evaluate the reproducibility of the CAJSA technique in healthy subjects and in patients with rheumatoid arthritis.
Methods: All joint space distance (JSD) measurements were performed using CAJSA-technology at the metacarpophalangeal articulation based on conventional and digital hand radiographs.
Results: I. Hand rotation showed an effect on the reproducibility with CV ranging between 0.39% (angulation 1°) and 1.66% (angulation 19°). II. Regarding the overall reproducibility of the CAJSA method, the intra-radiograph reproducibility of JSD was calculated with CV=0.54% for conventional images and CV=0.38% for digital images. The inter-radiograph reproducibility error was observed with CV=0.66% (conventional images) and CV=0.63% (digital images). III. The study revealed a reproducibility for CAJSA measurements in RA ranging between 0.37% (JSD-MCP ring finger; van der Heijde-modification of the Sharp method score 1) to 1.37% (JSD-MCP index finger; van der Heijde-modification of the Sharp method score 3).
Conclusion: CAJSA measures JSD at the metacarpophalangeal articulation with high reproducibility in healthy subjects and in patients with differing stages of rheumatoid arthritis. Additional findings show that hand rotation during X-ray imaging has an impact on the CAJSA measurements; thus avoiding acquisition of hand radiographs with a rotation error of more than 15 degrees can be recommended.
Copyright © 2011 Société française de rhumatologie. Published by Elsevier SAS. All rights reserved.