Background: A multi-center study was conducted to evaluate the Bio-Rad GS HIV Combo Ag/Ab EIA, a 4th generation HIV-1/HIV-2 assay for the simultaneous detection of HIV p24 antigen and antibodies to HIV-1 (groups M and O) and HIV-2 in human serum or plasma in adult and pediatric populations.
Objectives: The objectives of the study were to assess assay performance for the detection of acute HIV infections; sensitivity in known HIV positive samples; percent agreement with HIV status; specificity in low and high risk individuals of unknown HIV status; and to compare assay performance to a 3rd generation HIV assay.
Study design: The evaluation included testing 9150 samples at four U.S. clinical trial sites, using three kit lots. Unlinked samples were from routine testing, repositories or purchased from vendors.
Results: GS HIV Combo Ag/Ab EIA detection in samples from individuals in two separate populations with acute HIV infection was 95.2% (20/21) and 86.4% (38/44). Sensitivity was 100% (1603/1603) in known antibody positive [HIV-1 Groups M and O, and HIV-2] samples. HIV p24 antigen detection was 100% (53/53) in HIV-1 culture supernatants. HIV-1 seroconversion panel detection improved by a range of 0-20 days compared to a 3rd generation HIV test. Specificity was 99.9% (5989/5996) in low risk, 99.9% (959/960) in high risk and 100% (100/100) in pediatric populations.
Conclusion: The GS HIV Combo Ag/Ab EIA significantly reduced the diagnostic window when compared to the 3rd generation screening assay, enabling earlier diagnosis of HIV infection. The performance parameters of the Bio-Rad GS HIV Combo Ag/Ab EIA are well suited for use in HIV diagnostic settings.
Copyright © 2011 Elsevier B.V. All rights reserved.