Ultrasonic integrated backscatter two-dimensional imaging: evaluation of M-mode guided acquisition and immediate analysis in 55 consecutive patients

J Am Soc Echocardiogr. 1990 Jul-Aug;3(4):255-65. doi: 10.1016/s0894-7317(14)80308-8.

Abstract

We have shown previously that cardiac cycle-dependent integrated backscatter characterizes the physical state of myocardium in patients with ischemic heart disease and cardiomyopathy. In the present study the clinical applicability of M-mode guided two-dimensional integrated backscatter imaging was defined in evaluation of 55 nonselected patients. The mean amplitude of cyclic variation of integrated backscatter in normal segments (long-axis view) was as follows: basal septum, 4.2 +/- 1.3 dB (mean +/- SD; n = 27), mid-septum, 4.5 +/- 1.0 dB (n = 26), basal posterior, 4.8 +/- 1.0 dB (n = 30), and mid-posterior, 4.8 +/- 1.2 decibels (n = 27). The respective mean delay values (R wave to nadir) were as follows: 0.89 +/- 0.09, 0.84 +/- 0.09, 0.86 +/- 0.09, and 0.85 +/- 0.12. At least one cardiac cycle could be analyzed fully in 62% of patients. Limitations included technically difficult two-dimensional echocardiography, inadequate M-line orientation, technically remediable errors, or poor quality integrated backscatter images. In abnormal segments (n = 13) cyclic variation was reduced and delay was prolonged (1.2 +/- 1.1 dB and 1.21 +/- 1.1, respectively). Intraobserver and interobserver variability for amplitude measurements were modest, with respective correlation coefficients of r = 0.93; r = 0.72. The findings demonstrate that M-mode--assisted integrated backscatter is a practical approach for characterization of regional myocardial properties promptly and at the bedside in a large portion of patients with cardiac disease.

Publication types

  • Research Support, U.S. Gov't, P.H.S.

MeSH terms

  • Echocardiography* / methods*
  • Heart Diseases / diagnostic imaging*
  • Humans
  • Image Processing, Computer-Assisted*
  • Observer Variation