Chemical exchange saturation transfer magnetic resonance imaging of human knee cartilage at 3 T and 7 T

Magn Reson Med. 2012 Aug;68(2):588-94. doi: 10.1002/mrm.23250. Epub 2011 Dec 27.

Abstract

The sensitivity of chemical exchange saturation transfer (CEST) on glycosaminoglycans (GAGs) in human knee cartilage (gagCEST) in vivo was evaluated at 3 and 7 T field strengths. Calculated gagCEST values without accounting for B(0) inhomogeneity (~0.6 ppm) were >20%. After B(0) inhomogeneity correction, calculated gagCEST values were negligible at 3 T and ~6% at 7 T. These results suggest that accurate B(0) correction is a prerequisite for observing reliable gagCEST. Results obtained with varying saturation pulse durations and amplitudes as well as the consistency between numerical simulations and our experimental results indicate that the negligible gagCEST observed at 3 T is due to direct saturation effects and fast exchange rate. As GAG loss from cartilage is expected to result in a further reduction in gagCEST, gagCEST method is not expected to be clinically useful at 3 T. At high fields such as 7 T, this method holds promise as a viable clinical technique.