Background and purpose: In contrast to early migration, the long-term migration of hydroxyapatite- (HA-) coated tibial components in TKA has been scantily reported. This randomized controlled trial investigated the long-term migration measured by radiostereometric analysis (RSA) of HA-coated, uncoated, and cemented tibial components in TKA.
Patients and methods: 68 knees were randomized to HA-coated (n = 24), uncoated (n = 20), and cemented (n = 24) components. All knees were prospectively followed for 11-16 years, or until death or revision. RSA was used to evaluate migration at yearly intervals. Clinical and radiographic evaluation was according to the Knee Society system. A generalized linear mixed model (GLMM, adjusted for age, sex, diagnosis, revisions, and BMI) was used to take into account the repeated-measurement design.
Results: The present study involved 742 RSA analyses. The mean migration at 10 years was 1.66 mm for HA, 2.25 mm for uncoated and 0.79 mm for the cemented group (p < 0.001). The reduction of migration by HA as compared to uncoated components was most pronounced for subsidence and external rotation. 3 tibial components were revised for aseptic loosening (2 uncoated and 1 cemented), 3 for septic loosening (2 uncoated and 1 cemented), and 1 for instability (HA-coated). 2 of these cases were revised for secondary loosening after a period of stability: 1 case of osteolysis and 1 case of late infection. There were no statistically significant differences between the fixation groups regarding clinical or radiographic scores.
Interpretation: HA reduces migration of uncemented tibial components. This beneficial effect lasts for more than 10 years. Cemented components showed the lowest migration. Longitudinal follow-up of TKA with RSA allows early detection of secondary loosening.