The laboratory diagnosis of antiphospholipid antibody syndrome (APS) requires the demonstration of antiphospholipid antibodies (aPL) by lupus anticoagulant (LAC) measured through coagulation assays, anticardiolipin IgG or IgM antibodies (aCL) and/or anti-β2-glycoprotein I IgG or IgM antibodies (anti-β2-GPI), usually detected by ELISA. In this study we tested aCL by a new automated system using the chemiluminescence principle. Our results showed that, while almost all the sera from APS patients, positive for IgG aCL and anti-β2-GPI by ELISA, were also positive for IgG aCl by chemiluminescence, only 30.13% of patients without clinical manifestations of APS, but positive for aCL and persistently negative for anti-β2-GPI (by ELISA) and LA, confirmed the positive test by chemiluminescence. This difference was highly significant (p<0.0001). Interestingly, this test also prompted to identify 20% of patients positive for LA, but persistently negative for both aCL and anti-β2-GPI IgG (ELISA). Thus, the new technology of automated chemiluminescence assay for measuring aPL may represent an useful tool to identify "true" APS patients.
Copyright © 2012 Elsevier B.V. All rights reserved.