Twelve-month results of a paclitaxel releasing balloon in patients presenting with in-stent restenosis First-in-Man (PEPPER) trial

Cardiovasc Revasc Med. 2012 Sep-Oct;13(5):260-4. doi: 10.1016/j.carrev.2012.06.002. Epub 2012 Aug 4.

Abstract

Background: Coronary in-stent restenosis (ISR) continues to be a therapeutic challenge especially after drug eluting stent (DES) implantation. We studied patients with ISR to investigate safety and efficacy of a novel drug coated balloon (DCB) incorporating paclitaxel into a microcrystalline structure by applying the inert excipient butyryltri-n-hexyl citrate (BTHC) in a prospective First-in-Man trial.

Methods and materials: Eighty-one patients were enrolled at 9 European sites, thereof 43 (53.1%) presenting with bare metal stent (BMS)-ISR and 38 (46.9%) with DES-ISR. The primary study endpoint was in-stent late lumen loss (LLL) independently assessed by a quantitative coronary angiography laboratory at 6 months. A secondary endpoint was major adverse cardiac events (MACE), a composite of cardiac death, non-fatal myocardial infarction, clinically driven target vessel revascularization after 6 and 12 months.

Results: At 6 months, overall LLL was 0.07±0.31 mm showing differences in BMS-ISR and DES-ISR treatment (-0.05±0.28 mm vs. 0.19±0.29 mm, respectively, P=.001). Overall MACE rates at 6 and 12 months were 6.5% and 11.8%. At the 12-month follow-up, one myocardial infarction, and no cardiac death nor stent thrombosis had occurred.

Conclusion: Application of a novel paclitaxel coated balloon using BTHC as an excipient in patients with ISR is safe and results in very low LLL, revascularization- and MACE-rates at follow-up. (ClinicalTrials.gov: NCT00961181).

Publication types

  • Clinical Trial
  • Multicenter Study
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Aged
  • Angioplasty, Balloon, Coronary / adverse effects
  • Angioplasty, Balloon, Coronary / instrumentation*
  • Angioplasty, Balloon, Coronary / mortality
  • Butyrates / chemistry
  • Cardiovascular Agents / administration & dosage*
  • Coated Materials, Biocompatible*
  • Coronary Angiography
  • Coronary Restenosis / diagnostic imaging
  • Coronary Restenosis / etiology
  • Coronary Restenosis / mortality
  • Coronary Restenosis / therapy*
  • Coronary Stenosis / mortality
  • Coronary Stenosis / therapy*
  • Drug-Eluting Stents
  • Europe
  • Excipients / chemistry
  • Feasibility Studies
  • Female
  • Humans
  • Male
  • Metals
  • Middle Aged
  • Myocardial Infarction / etiology
  • Paclitaxel / administration & dosage*
  • Prospective Studies
  • Prosthesis Design
  • Stents*
  • Time Factors
  • Treatment Outcome

Substances

  • Butyrates
  • Cardiovascular Agents
  • Coated Materials, Biocompatible
  • Excipients
  • Metals
  • butyryl-n-trihexylcitrate
  • Paclitaxel

Associated data

  • ClinicalTrials.gov/NCT00961181