High-frequency rapid B-mode ultrasound imaging for real-time monitoring of lesion formation and gas body activity during high-intensity focused ultrasound ablation

IEEE Trans Ultrason Ferroelectr Freq Control. 2012 Aug;59(8):1687-99. doi: 10.1109/TUFFC.2012.2374.

Abstract

The goal of this study was to examine the ability of high-frame-rate, high-resolution imaging to monitor tissue necrosis and gas-body activities formed during high-intensity focused ultrasound (HIFU) application. Ex vivo porcine cardiac tissue specimens (n = 24) were treated with HIFU exposure (4.33 MHz, 77 to 130 Hz pulse repetition frequency (PRF), 25 to 50% duty cycle, 0.2 to 1 s, 2600 W/cm(2)). RF data from B-mode ultrasound imaging were obtained before, during, and after HIFU exposure at a frame rate ranging from 77 to 130 Hz using an ultrasound imaging system with a center frequency of 55 MHz. The time history of changes in the integrated backscatter (IBS), calibrated spectral parameters, and echo-decorrelation parameters of the RF data were assessed for lesion identification by comparison against gross sections. Temporal maximum IBS with +12 dB threshold achieved the best identification with a receiver-operating characteristic (ROC) curve area of 0.96. Frame-to-frame echo decorrelation identified and tracked transient gas-body activities. Macroscopic (millimeter-sized) cavities formed when the estimated initial expansion rate of gas bodies (rate of expansion in lateral-to-beam direction) crossed 0.8 mm/s. Together, these assessments provide a method for monitoring spatiotemporal evolution of lesion and gas-body activity and for predicting macroscopic cavity formation.

Publication types

  • Research Support, N.I.H., Extramural
  • Research Support, Non-U.S. Gov't

MeSH terms

  • Animals
  • Echocardiography / instrumentation
  • Echocardiography / methods*
  • Gases / analysis*
  • Gases / chemistry
  • High-Intensity Focused Ultrasound Ablation / instrumentation
  • High-Intensity Focused Ultrasound Ablation / methods*
  • High-Intensity Focused Ultrasound Ablation / standards
  • Image Processing, Computer-Assisted
  • Myocardium / pathology*
  • Necrosis / diagnostic imaging
  • Necrosis / pathology
  • ROC Curve
  • Swine
  • Transducers

Substances

  • Gases