Dental models made with an intraoral scanner: a validation study

Am J Orthod Dentofacial Orthop. 2012 Sep;142(3):308-13. doi: 10.1016/j.ajodo.2012.03.031.

Abstract

Introduction: Our objectives were to determine the validity and reproducibility of measurements on stereolithographic models and 3-dimensional digital dental models made with an intraoral scanner.

Methods: Ten dry human skulls were scanned; from the scans, stereolithographic models and digital models were made. Two observers measured transversal distances, mesiodistal tooth widths, and arch segments on the skulls and the stereolithographic and digital models. All measurements were repeated 4 times. Arch length discrepancy and tooth size discrepancy were calculated. Statistical analysis was performed by using paired t tests.

Results: For the measurements on the stereolithographic and digital models, statistically significant differences were found. However, these differences were considered to be clinically insignificant. Digital models had fewer statistically significant differences and generally the smallest duplicate measurement errors compared with the stereolithographic models.

Conclusions: Stereolithographic and digital models made with an intraoral scanner are a valid and reproducible method for measuring distances in a dentition.

Publication types

  • Validation Study

MeSH terms

  • Cadaver
  • Chi-Square Distribution
  • Computer Simulation*
  • Humans
  • Image Processing, Computer-Assisted*
  • Imaging, Three-Dimensional / methods
  • Models, Dental*
  • Odontometry / methods
  • Photography, Dental / instrumentation
  • Photography, Dental / methods*
  • Reproducibility of Results
  • Software Validation