Background and purpose: Effects of blood flow modification by flow diverters are observed to lead often to aneurysm thrombosis and reverse remodeling. For this process, to further understand the potential roles of intra-aneurysmal blood pressure changes and aneurysm morphologies, 23 patients were studied by numeric simulation.
Materials and methods: 3D imaging of aneurysms of different sizes and shapes, all located at the supraclinoid segment of the ICA (n=23), was prepared for CFD simulations. Hemodynamic variables were calculated for conditions before and after virtual FD implantation, reconstituting a vessel wall scaffold across the aneurysm neck. WSS, velocity, residence time, turnover time, and intra-aneurysmal pressure were assessed statistically.
Results: After placement of FDs, significant reductions inside the aneurysm were observed for most hemodynamic variables (P<.01) except mean intra-aneurysmal pressures. For minimum/maximum intra-aneurysmal pressure values, small but significant changes were found; however, they were considered too small to be of relevance.
Conclusions: Calculations in 23 cases did not reveal significant intra-aneurysmal mean or peak pressure changes, indicating a minor role of pressure changes in the rare event of secondary ruptures after FD use. Other hemodynamic variables (WSS and velocity) exhibited more significant changes, indicating their role in intra-aneurysmal thrombus formation. Size-dependent, significantly higher reduction in WSS (P=.069) and velocity (P=.013) was observed in small aneurysms compared with larger ones. When it came to shape, there were significantly higher reductions in WSS (P=.055) and velocity (P=.065) and a significantly higher increase in turnover time in fusiform aneurysms compared with saccular aneurysms.