Long-term outcomes after transcatheter aortic valve implantation: insights on prognostic factors and valve durability from the Canadian multicenter experience

J Am Coll Cardiol. 2012 Nov 6;60(19):1864-75. doi: 10.1016/j.jacc.2012.08.960. Epub 2012 Oct 10.

Abstract

Objectives: This study sought to evaluate the long-term outcomes after transcatheter aortic valve implantation (TAVI) in the Multicenter Canadian Experience study, with special focus on the causes and predictors of late mortality and valve durability.

Background: Very few data exist on the long-term outcomes associated with TAVI.

Methods: This was a multicenter study including 339 patients considered to be nonoperable or at very high surgical risk (mean age: 81 ± 8 years; Society of Thoracic Surgeons score: 9.8 ± 6.4%) who underwent TAVI with a balloon-expandable Edwards valve (transfemoral: 48%, transapical: 52%). Follow-up was available in 99% of the patients, and serial echocardiographic exams were evaluated in a central echocardiography core laboratory.

Results: At a mean follow-up of 42 ± 15 months 188 patients (55.5%) had died. The causes of late death (152 patients) were noncardiac (59.2%), cardiac (23.0%), and unknown (17.8%). The predictors of late mortality were chronic obstructive pulmonary disease (hazard ratio [HR]: 2.18, 95% confidence interval [CI]: 1.53 to 3.11), chronic kidney disease (HR: 1.08 for each decrease of 10 ml/min in estimated glomerular filtration rate, 95% CI: 1.01 to 1.19), chronic atrial fibrillation (HR: 1.44, 95% CI: 1.02 to 2.03), and frailty (HR: 1.52, 95% CI: 1.07 to 2.17). A mild nonclinically significant decrease in valve area occurred at 2-year follow-up (p < 0.01), but no further reduction in valve area was observed up to 4-year follow-up. No changes in residual aortic regurgitation and no cases of structural valve failure were observed during the follow-up period.

Conclusions: Approximately one-half of the patients who underwent TAVI because of a high or prohibitive surgical risk profile had died at a mean follow-up of 3.5 years. Late mortality was due to noncardiac comorbidities in more than one-half of patients. No clinically significant deterioration in valve function was observed throughout the follow-up period.

Publication types

  • Multicenter Study

MeSH terms

  • Aged
  • Aged, 80 and over
  • Canada / epidemiology
  • Cardiac Catheterization / mortality
  • Cardiac Catheterization / trends*
  • Equipment Design / mortality
  • Equipment Design / trends*
  • Female
  • Follow-Up Studies
  • Heart Valve Diseases / diagnostic imaging*
  • Heart Valve Diseases / mortality
  • Heart Valve Diseases / surgery*
  • Heart Valve Prosthesis Implantation / mortality
  • Heart Valve Prosthesis Implantation / trends*
  • Humans
  • Male
  • Prognosis
  • Prospective Studies
  • Survival Rate / trends
  • Time Factors
  • Treatment Outcome
  • Ultrasonography