Objective: This study evaluated the mechanical characteristics and stress distribution of the hybrid technique fixation of the sagittal split ramus osteotomy.
Study design: In the mechanical test, 10 polyurethane replicas of human hemimandibles of each group were submitted to linear loading test. For the photoelastic evaluation, 3 hemimandible replicas of photoelastic resin were subjected to photoelastic analysis. In the finite element analysis, 3 computer models simulated the displacement and the results of maximum principal stress were analyzed.
Results: The results of this study demonstrated that the fixation technique with 3 bicortical screws presented better mechanical resistance and stress distribution pattern when compared with the hybrid technique that, on the other hand, presents better results in comparison with a miniplate and monocortical screws.
Conclusions: The results suggest that the hybrid technique increased the resistance and improved stress distribution of miniplate/monocortical screw fixation, maintaining most of the advantages of this technique.