Cardiac device-related endocarditis: 31-Years' experience

J Cardiol. 2013 Feb;61(2):175-80. doi: 10.1016/j.jjcc.2012.08.019. Epub 2012 Nov 17.

Abstract

Background: Cardiac device-related endocarditis (CDE) is a major complication of the implantation of a pacemaker and defibrillator. The experience in a single high-volume tertiary center is reported.

Methods: Thirty one years (1980-2011) of cases of CDE were analyzed retrospectively and compared to overall insertion data; the clinical course and management strategies of these patients have been reviewed.

Results: A total of 23 cases (16 male, median age 72 years) were identified, 20 of these cases were determined at our institution where 5287 procedures were performed (endocarditis rate 0.38%). Thirteen patients were determined to have a cardiac device pocket infection. Infection in 7 cases (30%) was caused by lead(s). However, in 16 cases (70%) both leads and the pocket of devices were the reason of infection. Median time was 13.5 months for presentation. Patients who had undergone the last procedure within 6 months were admitted earlier than those with longer post procedure time (p<0.05). Transesophageal echocardiography demonstrated lead vegetations in 13 of the 16 cases (81%). Organisms were identified in 18 cases (78%)-78% Staphylococci (56% Staphylococcus aureus). Leads of the device were removed in 17 cases (74%); seven cases by percutaneous simple traction and 10 cases by sternotomy. Six major complications attributable to device-related endocarditis were observed: four deaths (mortality 17.4%); one splenic abscess requiring splenectomy; and one septic pulmonary embolism; median follow-up 49 months.

Conclusion: A CDE endocarditis rate of 0.38% was demonstrated. It remains a rare but potentially lethal complication of device implantation.

MeSH terms

  • Adult
  • Aged
  • Aged, 80 and over
  • Anti-Bacterial Agents / therapeutic use
  • Bacteremia / etiology
  • Brucella
  • Defibrillators, Implantable / adverse effects
  • Device Removal / adverse effects
  • Device Removal / methods
  • Echocardiography / methods
  • Endocarditis, Bacterial / etiology
  • Enterococcus faecalis
  • Female
  • Humans
  • Male
  • Middle Aged
  • Pacemaker, Artificial / adverse effects
  • Prosthesis-Related Infections / etiology*
  • Pseudomonas aeruginosa
  • Retrospective Studies
  • Staphylococcus
  • Treatment Outcome

Substances

  • Anti-Bacterial Agents